A metrological calibration facility for particle number counters used in emission measurements
Abstract. A metrological calibration facility for particle number counters (PNCs) has been developed at the Physikalisch-Technische Bundesanstalt (PTB) as a primary reference standard for this metric required in the quantification of solid soot emission as described in national and European regulatory frameworks. This study describes the calibration of a Condensation Particle Counter (CPC) using PTB’s primary aerosol number standard, a Faraday Cup Aerosol Electrometer (FCAE). The calibration facility enables high accuracy determination of CPC counting efficiency for particles with mobility diameters between 17 and 200 nm, as well as linearity measurements of particle number concentrations ranging from 1 cm⁻³ to 20 000 cm⁻³. Soot aerosol particles were generated using a miniCAST 5303 and subsequently treating them in a Catalytic Stripper to remove volatile components before homogenisation in an integrated mixing unit. The aerosol flow was distributed via a ten-port manifold, allowing simultaneous evaluation of up to eight CPCs and a FCAE connected to two different ports simultaneously. The results demonstrate that the facility provides state-of-the-art metrological capabilities for CPC calibration, supporting improved reliability and comparability of particle number measurements in emission control applications.
- Preprint
(1321 KB) - Metadata XML
- BibTeX
- EndNote
Status: final response (author comments only)
- RC1: 'Comment on ar-2026-33', Anonymous Referee #1, 08 Oct 2026
-
RC2: 'Comment on ar-2026-33', Anonymous Referee #2, 12 Nov 2026
The authors discuss the use of a Faraday Cup Aerosol Electrometer as a reference measurement tool for particle counting at a calibration facility in Germany. Authors have significant amount of data and they have discussed various details in the manuscript but the manuscript needs major revisions to the current version before it can be considered for acceptance. A few major comments:
- Grammar needs to be checked thoroughly throughout the paper to ensure good quality for the reader without being annoyed.
- This type of work is not new but the manuscript only cites their own published articles or very few other ones. Authors need to properly credit other researchers who have made advancements in previous years wherever possible. This is a serious issue that needs to be addressed.
- Introduction section needs to be improved significantly.
- Description of figures not written properly (e.g. for figures 13-17, no discussion and interpretation of the figures provided)
- Write full forms of the acronyms at the first use in the manuscript, many acronyms were used without writing the details.
Other comments:
- line 9: it says "a" CPC but the work is focusing on multiple counters
- line 26: Write the full form of "SI-traceable".
- line 53: May consider using the term "better" in place of "lower"
- line 59: Please fix the language (does it mean that the generator was running 24/7 during that period?)
- line 66: "a" not needed before +-30mL/min
- line 69-72: Are the authors discussing the PTB facility or is it a general comment?
- line 76: improve sentence
- Figure 3: Please improve the schematic. in this section, authors also need to refer to the other sections later in the manuscript e.g. multiple charging issue, size and operation of this facility. Adding an actual image of the system may be beneficial.
- line 93: sheath to aerosol flow ratio should be 10:1, it's written the opposite way
- line 104: Consider using the term "maintaining" in place of "to maintain"
- line 109: Consider using the term "custom-designed" or "custom-made" in place of "self-made"
- Figure 5: again, a short description of the dimensions missing
- line 120-125: Why do you need such high flow rates? Authors should discuss this a little about it.
- Figure 6: Has this design not been published earlier? If yes, Is this image from the patent?? Please provide details and cite accordingly.
- line 135: Consider writing "To achieve the desired particle number concentration, the inlet flow to the MFS was adjusted in a range..."
- line 139: remove "were"
- line 142, figure 3: If ports 1a and 1b had the highest bias then why is the rationale behind having the electrometer on these two ports? Why not, use it on 2a or other ports with lower bias? Shouldn't the reference measurement be the best possible and stable measurement?
- line 164-166: "MCC"? I assume the authors are referring to multiple charge correction? Please improve this section. Readers need to be able to understand the any graphs in the manuscript without referring to other documents all the time. Please describe the stepping method in brief and mention what 1U, 2U, and 3U mean.
- line 209: it should be "PTB's"; please fix grammar as well
Citation: https://doi.org/10.5194/ar-2026-33-RC2
Viewed
| HTML | XML | Total | BibTeX | EndNote | |
|---|---|---|---|---|---|
| 361 | 167 | 45 | 573 | 50 | 50 |
- HTML: 361
- PDF: 167
- XML: 45
- Total: 573
- BibTeX: 50
- EndNote: 50
Viewed (geographical distribution)
| Country | # | Views | % |
|---|
| Total: | 0 |
| HTML: | 0 |
| PDF: | 0 |
| XML: | 0 |
- 1
This manuscript describes PTB's calibration facility for particle counters used in emission control. Although I greatly appreciate the effort that has been put into setting up and validating the system, the manuscript merely describes a method that has been standardised within ISO 27891 for almost 15 years. As the authors acknowledge, several other national metrology institutes and calibration laboratories around the world have developed and validated similar setups and at least two international intercomparisons have taken place (Högström et al., 2014) (Brown et al., 2023).
After reading the reports of the intercomparisons (Högström et al., 2014; Brown et al., 2023), it became clear to me that several other metrology institutes in Europe and Asia have smaller measurement uncertainties and better agreement with the reference value than PTB. Surprisingly, this issue has not been discussed in the manuscript.
Moreover, a quick literature search revealed that there are at least two other papers describing methods to calibrate CPCs and other particle counters down to 1 cm-3 in a traceable manner (Sakurai et al., Metrologia 60(5), 2023, 10.1088/1681-7575/acf5f1; Owen et al. Aerosol Science & Technology 46, 2012, https://doi.org/10.1080/02786826.2011.637998) which the authors have not cited. Note that the paper by Owen et al. dates back to 2012, showing that NIST had developed a CPC calibration method based on FCAE and dilution units already 15 years ago.
The suitability of miniCAST generators as a source of combustion particles has been examined and confirmed in numerous studies. I cannot cite here all the papers, but please see, for instance, Ess et al. (Aerosol Science and Technology, 53(1), 2019, https://doi.org/10.1080/02786826.2018.1536818) and references therein. Therefore, there is nothing novel in this regard either.
So much work has been done in the last 15-20 years regarding aerosol/combustion generators, design and validation of multiple-flow splitters, calibration of particle counters (CPCs and diffusion chargers) and electrometers, yet the authors of this manuscript cite a very small number of papers, the majority of which has been co-authored by them. The most striking example is when the authors state that "The Faraday Cup Aerosol Electrometer (FCAE) has been established as the primary standard for PN determination (Molzahn et al., 2025)". It was not PTB (Molzahn et al.) who established FCAE as a primary standard in 2025. FCAEs have been used for more than 15 years as primary standards for number concentration all around the world!
To summarise, I respect the efforts invested in validating the PTB setup. However, this setup does not go beyond the-state-of-the-art in any way. The calibration method described in this manuscript was standardised within ISO 21789 more than 15 years ago and many other institutes have developed similar setups. I am astonished that the authors do not acknowledge previous work done in their field. I also have the impression that this manuscript serves more like a "marketing" strategy to advertise PTB calibration services rather than as a means to communicate new scientific knowledge that helps advance aerosol science. For these reasons, I do not recommend publication of this manuscript.