Size distribution and particle morphology of analytes dried through the evaporative light scattering detector – Part 1

Frederick Bertani, Joshua Hassim, and Simone Hochgreb

Related authors

Simulation of aerosol transport, evaporation and scattering in the Evaporative Light Scattering Detector – Part 2
Frederick Bertani, Joshua Hassim, and Simone Hochgreb
Aerosol Research, 4, 325–344, https://doi.org/10.5194/ar-4-325-2026,https://doi.org/10.5194/ar-4-325-2026, 2026
Short summary

Cited articles

Adelantado, C., Rodríguez-Fariñas, N., Rodríguez Martín-Doimeadios, R. C., Zougagh, M., and Ángel Ríos: Analysis of silica nanoparticles by capillary electrophoresis coupled to an evaporative light scattering detector, Anal. Chim. Acta, 923, 82–88, https://doi.org/10.1016/j.aca.2016.03.055, 2016. a
Aliseda, A., Hopfinger, E. J., Lasheras, J. C., Kremer, D. M., Berchielli, A., and Connolly, E. K.: Atomization of viscous and non-Newtonian liquids by a coaxial, high-speed gas jet. Experiments and droplet size modeling, Int. J. Multiphas. Flow, 34, 161–175, 2008. a, b, c, d
Bertani, F.: Improving Evaporative Light Scattering Detector performance using Experiments and Modelling, PhD thesis, Apollo – University of Cambridge Repository, https://doi.org/10.17863/CAM.116176, 2024. a, b
Bertani, F., Hassim, J., and Hochgreb, S.: Simulation of aerosol transport, evaporation and scattering in the Evaporative Light Scattering Detector – Part 2, Aerosol Res., 4, 325–344, https://doi.org/10.5194/ar-4-325-2026, 2026. a, b, c, d
Boborodea, A., Mirabella, F. M., and O'Donohue, S.: Polyolefin characterization in xylene by high-temperature gel permeation chromatography with a new evaporative light scattering detector, Chromatographia, 81, 419–424, 2018. a
Short summary
This paper presents the very first measurements of droplet and particle sizes and concentrations through the parts of an elastic light scatter detector. Measurements were made using two instruments, phase Doppler particle anemometry and an aerodynamic aerosol classifier, to resolve the size distribution of volatile particles upstream and dried particles downstream. The final measurements are used in a companion paper comparison with simulations.